Enforcement Report - Week of December 19, 2012
-
Back to Previous Page
Event Detail
| Event ID | 63312 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Roche Molecular Systems, Inc. |
| City | Branchburg |
| State | NJ |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-06-25 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Nationwide Distribution |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| AMPLILINK software, For use with COBAS® AmpliPrep instrument, COBAS® TaqMan® analyzer, COBAS® TaqMan® 48 analyzer, COBAS® AMPLICOR® analyzer, and Cobas p 630 instrument AMPLILINK software is used to integrate all instruments and analyzers offering a single user interface for PCR testing. | AMPLILINK version 3.3 and 3.2 | Class II | Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . | 884 units | Z-0484-2013 |
| Roche Applied Science MagNA Pure LC Version 3.0. The MagNA Pure LC (version 3.0) is a sample preparation instrument that can be used with the LightCycler instrument for automated isolation of nucleic acids in various types of biological materials. | MagNA Pure LC version 3.0 | Class II | Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . | 816 units | Z-0485-2013 |
| Roche MagNA Pure 96 System Version 1.0 (Software Version 2.0) for in vitro diagnostic use. The MagNA Pure 96 System (versions 1.0 and 2.0) is a sample preparation instrument for automated isolation of nucleic acids in various types of biological materials. This is a general purpose instrument. There are no corresponding assays. | Roche MagNA Pure 96 System Version 1.0 (Software Version 2.0) | Class II | Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . | 13 units | Z-0486-2013 |
| Roche Molecular Biochemicals LightCycler 1.2 with software version 3.5. LightCycler 1.2 (software version 3.5) is a Rapid PCR analyzer with real-time detection and/or quantification of a target NA (nucleic acid), as well as post-PCR analysis of the amplified NA by melting curve analysis | Roche Molecular Biochemicals LightCycler 1.2 with software version 3.5. | Class II | Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . | 3,772 units | Z-0487-2013 |
| Roche LightCycler 2.0 Instrument with software version 4.1. LightCycler 2.0 Instrument (software version 4.1) is a Rapid PCR analyzer with real-time detection and/or quantification of a target NA (nucleic acid), as well as post-PCR analysis of the amplified NA by melting curve analysis. | Roche LightCycler 2.0 Instrument with software version 4.1. | Class II | Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . | 1,213 units | Z-0488-2013 |
| cobas® 8000 modular analyzer series The COBAS 8000 data manager modular analyzer series (all versions) is a is a fully automated, random-access, software- controlled system for immunoassay and photometric analysis intended for qualitative and quantitative in-vitro determinations using a wide variety of tests. | cobas® 8000 modular analyzer series | Class II | Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . | 77 units | Z-0489-2013 |
| cobas® Integra 400/400 Plus (Software version 3.4) The COBAS INTEGRA 400/400 plus system (software version 3.4) is a fully automated, computerized chemistry analyzer. It is intended for in vitro quantitative or qualitative determination of a wide range of analytes in various body fluids. | cobas® Integra 400/400 Plus (Software version 3.4) | Class II | Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . | 855 units. | Z-0490-2013 |
| DataInnovations (Instrument Manager) (version 8.04) interfaces middleware between larger Clinical Chemistry / Immunology instrument systems and LIS systems. | Instrument Manager version 8.04. | Class II | Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . | 1,015 units | Z-0491-2013 |
| DataCare GM DataCara GM is an information management system that provides data management functionality for managing blood glucose testing data. | DataCare GM | Class II | Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . | 479 units | Z-0492-2013 |
| DataCare POC is an information management system that provides data management functionality for managing point-of -care testing data. | DataCare POC | Class II | Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a "pre authentication vulnerability" which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ . | 126 units | Z-0493-2013 |
-