Enforcement Report - Week of December 19, 2012

  • Print
  • Share
  • E-mail
-

Event Detail

Event ID 63518
Product Type Devices
Status Ongoing
Recalling Firm Hospira Inc.
City Lake Forest
State IL
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-10-26
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution - USA (nationwide) including Puerto Rico, and the countries of Australia, Bahrain, Belgium, Brazil, Canada, Chile, Colombia, Denmark, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Zealand, Oman, Pakistan, Philippines, Poland, Portugal, Saudi Arabia, Singapore, Lavenia, South Africa, Spain, Switzerland, Taiwan, Turkey, United Kingdom, United Arab Emirates and Vietnam.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Plum A+ Single Channel Infusion Pumps with Plug-and-Play Modules that provide wired Ethernet and wireless local area networking; Hospira, Inc., Lake Forest, IL 60045; pump list number 11971, module list numbers 12101, 12102 and 12393; pump list number 11973, module list number 12380; pump list number 12391, module list numbers 12097 and 12680 Product Usage: The Plum A+ Infusion System is a dual-line volumetric infusion pump that provides precise delivery of multiple i.v. therapies across the general spectrum of clinical care, including ICU/CCU, OR, PACU, ER, Med/Surg, Pediatrics, Outpatient clinics and Home care pump list number 11971, module list numbers 12101, 12102 and 12393; pump list number 11973, module list number 12380; pump list number 12391, module list numbers 12097 and 12680; all serial numbers Class II The volume control knob on some Plum A+ single channel infusers (located on the back of the infuser) may not function as described in the System Operating Manual - the direction for Loud and Quiet may be reversed. The manual indicates a clockwise rotation of the knob decreases the volume; however, on some of the devices the alarm volume increases when the knob is turned clockwise. Z-0481-2013
Plum A+ Hyperbaric Single Channel Infusion Pumps with Hospira MedNet Software and a Connectivity Engine peripheral module that provides wired Ethernet and wireless local networking, allowing the MedNet software to download drug libraries to the infuser and enable the auto-programming feature; Hospira, Inc., Lake Forest, IL 60045; pump list number 11005, module list number 11006 Product Usage: The Plum A+ Hyperbaric Infusion System is exclusively designed for hyperbaric therapy in monospace and multiplace chambers. pump list number 11005 and module list number 11006; all serial numbers Class II The volume control knob on some Plum A+ single channel infusers (located on the back of the infuser) may not function as described in the System Operating Manual - the direction for Loud and Quiet may be reversed. The manual indicates a clockwise rotation of the knob decreases the volume; however, on some of the devices the alarm volume increases when the knob is turned clockwise. Z-0482-2013
Plum A+ Single Channel Infusion Pumps with Hospira Mednet Software and a Connectivity Engine peripheral module that provides wired Ethernet and wireless local area networking, allowing the MedNet software to download drug libraries to the infuser and enable the auto-programming feature; Hospira, Inc., Lake Forest, IL 60045; pump list #20679 and 20792, module list number 20677. The Plum A+ Infusion System with Hospira MedNet Software and a Connectivity Engine peripheral module that provides wired Ethernet and wireless local area networking, allowing the MedNet software to download drug libraries to the infuser and enable the auto-programming feature. pump list numbers 20679 and 20792, and module list number 20677; all serial numbers Class II The volume control knob on some Plum A+ single channel infusers (located on the back of the infuser) may not function as described in the System Operating Manual - the direction for Loud and Quiet may be reversed. The manual indicates a clockwise rotation of the knob decreases the volume; however, on some of the devices the alarm volume increases when the knob is turned clockwise. Z-0483-2013
-
-