Event Detail

Event ID 63553
Product Type Devices
Status Ongoing
Recalling Firm Philips And Neusoft Medical Systems Co., Ltd.
City Shenyang, Liaoning
State
Country CN
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-10-09
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Nationwide Distribution including North Carolina, Ohio, Texas, and Puerto Rico.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321 The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously. NeuViz 16 system Serial numbers: N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120022, N16E120020, N16E120023, N16E120024. Neuviz Dual Serial numbers: 400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002, NDHR080003, NDHR080004, NDHR080007, NDHR080012, NDHR080021, NDHR090003, NDHR090012, and NDHR090013. Class II The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners. 20 units of NeuViz 16 CT Scanner Systems and 25 units of NeuViz Dual series Z-0466-2013
-
-