Enforcement Report - Week of December 5, 2012
-
Back to Previous Page
Event Detail
| Event ID | 63583 |
|---|---|
| Product Type | Devices |
| Status | Terminated |
| Recalling Firm | Metrex Research, LLC. |
| City | Romulus |
| State | MI |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2011-09-20 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide distribution: USA (nationwide) and countries of: Australia, Canada, and Europe. |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| Pentron Clinical Metrex Research; Pinnacle Cure Sleeve, a protective barrier The intended use of these devices is to protect both patient and healthcare worker from transfer of microorganisms, body fluids and particulate matter. | Part Number Lot Number 4511 201008270110 | Class II | The firm initiated the recall due to a labeling error. The boxes labeled as Pinnacle Cure Sleeve 4511, which fits 11mm light guides, actually contain Pinnacle Cure Sleeves that fit 7mm or 8mm light guides (part number 4500). | 3,348 units | Z-0444-2013 |
-