Event Detail

Event ID 63597
Product Type Devices
Status Ongoing
Recalling Firm GE Healthcare, LLC
City Waukesha
State WI
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-09-20
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Nationwide Distribution

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
GE Healthcare Brivo XR 285 Mobile general purpose radiographic imaging of the human head and body. part numbers:5555000, 5555000-2 Class II It was discovered GE Healthcare has become aware of a firmware issue on the mobile radiographic product, Brivo XR285amx, Optima XR200amx, Optima XR22Oamx and Optima XR200amx with Digital Upgrade, related to Technique Accuracy as mandated by the US Code of Federal Regulations (21CFR) and IEC 60601-2-54. 364 total units installed in US Z-0200-2013
GE Healthcare Optima 200amx Mobile general purpose radiographic imaging of the human head and body. part numbers: 5555000-3, 5555000-4 Class II It was discovered GE Healthcare has become aware of a firmware issue on the mobile radiographic product, Brivo XR285amx, Optima XR200amx, Optima XR22Oamx and Optima XR200amx with Digital Upgrade, related to Technique Accuracy as mandated by the US Code of Federal Regulations (21CFR) and IEC 60601-2-54. 364 total units installed in US Z-0199-2013
GE Healthcare Optima XR 220amx Mobile general purpose radiographic imaging of the human head and body. part numbers: 5555000-5, 5555000-6 Class II It was discovered GE Healthcare has become aware of a firmware issue on the mobile radiographic product, Brivo XR285amx, Optima XR200amx, Optima XR22Oamx and Optima XR200amx with Digital Upgrade, related to Technique Accuracy as mandated by the US Code of Federal Regulations (21CFR) and IEC 60601-2-54. 364 total units installed in US Z-0198-2013
GE Healthcare Optima XR 200amx with digital upgrade Mobile general purpose radiographic imaging of the human head and body. part number: 5421698 Class II It was discovered GE Healthcare has become aware of a firmware issue on the mobile radiographic product, Brivo XR285amx, Optima XR200amx, Optima XR22Oamx and Optima XR200amx with Digital Upgrade, related to Technique Accuracy as mandated by the US Code of Federal Regulations (21CFR) and IEC 60601-2-54. 364 total units installed in US Z-0197-2013
-
-