Enforcement Report - Week of January 9, 2013
-
Back to Previous Page
Event Detail
| Event ID | 63616 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Mindray DS USA, Inc. d.b.a. Mindray North America |
| City | Mahwah |
| State | NJ |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2011-05-12 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Nationwide Distribution including OH and GA |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| DPM Central Station Mindray DS USA, Inc. 800 MacArthur Boulevard, Mahwah, NJ 07430 The Central Monitoring Station (CMS) network transfers information between DOM Central Station and other network devices. It also allows information transfer between several CMS. Network connections consist of hardwired network cables and or WLAN connections. The DPM Central Station is used for remote monitor management, storing, printing, reviewing, or processing of information from net worked devices, and it is operated by medical personnel in hospitals or medical institutions. | 510 K K080192 P/N 300BF-PA200002 | Class II | There is a software anomaly with the DPM Central Monitoring System where the trend data of one patient maybe replaced with the trend data of a second patient. | two units | Z-0630-2013 |
-