Enforcement Report - Week of January 30, 2013
-
Back to Previous Page
Event Detail
| Event ID | 63617 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Roche Diagnostics Operations, Inc. |
| City | Indianapolis |
| State | IN |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-11-19 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Nationwide Distribution including AL, AR, AZ, CA, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY and GUAYNABO, PR. |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| Roche COBAS INTEGRA 400 and 400 plus Analyzer The Roche COBAS INTEGRA 400/400 plus Analyzer is a fully automated system for clinical chemistry analysis intended for the in vitro quantitative/qualitative determination of analytes in body fluids. | Part Numbers:28065047001 and 03245233001 | Class III | A software security issue with Oracles TNS-Listener component has been identified. TNS-Listener is supplied with the Roche COBAS INTEGRA 400/400 plus Analyzers. The Oracle TNS-Listener can be considered as an authentication and redirection component between the client and the database inside the Roche product. This means an attacker could register their own non-Roche database via remote access over the network to the Roche product and read and write database queries without authentication and therefore modify the content of a database. | 845 Units | Z-0696-2013 |
-