Enforcement Report - Week of December 12, 2012
-
Back to Previous Page
Event Detail
| Event ID | 63619 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Philips Healthcare Inc. |
| City | Andover |
| State | MA |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-10-30 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia, Austria, Belgium, Canada, Chile, Croatia, Czech Republic, Denmark, Egypt, Eastonia, Finland, France, Germany, India, Indonesia, Iraq, Italy, Japan, Latvia, Lithuania, Lybia, Malta, Mexico, Netherlands, New Zealnd, Norway, Oman, Polnad, Protugual, Quatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Soth Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, UAE and UK. |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| Philips Digital Diagnost Software Version R2.0.2 Product Usage: This system is used for making X-ray exposures for diagnostics | All Digital Diagnost systems with release 2.0.2 and 2.0.2 SP1 Models 712020, 712022 | Class II | With Digital Diagnost R2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a circle" appears in the lower left corner. This can be mistaken for a "Right patient side" marker, although this can appear on the left patient side. | 623 units | Z-0475-2013 |
-