Enforcement Report - Week of December 12, 2012
-
Back to Previous Page
Event Detail
| Event ID | 63646 |
|---|---|
| Product Type | Devices |
| Status | Completed |
| Recalling Firm | iwalk inc |
| City | Bedford |
| State | MA |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-10-26 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Nationwide Distribution including OH, NC, and NY. |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| BiOM Rechargeable Li Ion 22.2 V Battery and BiOM Lithium Battery Charger sold as an accessory to the BiOM lower leg system: Product Number: 2001249 BiOM Battery Product Number: 2001240 BiOM Charger | Product Code: 2001249 (BiOM Battery) ,Lot number:48831 Product Code: 2001240 (Charger), Lot number: 2973 | Class II | Battery may over-heat during charging and smoke | 21 Batteries; 9 Chargers | Z-0467-2013 |
-