Event Detail

Event ID 63671
Product Type Devices
Status Ongoing
Recalling Firm Philips Healthcare Inc.
City Andover
State MA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-10-30
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia, Austria, Canada, Chile, Czech republic, Estonia,France, Germany, Hungary, Inida, Ireland, Italy, Japan, Libyan Arab Jamahari, Mexico, Netherlands, New Zealand, Norway , Oman, Panama, Peru, Poland, Qatar, Reunion, Russia, Saudi Arabia, South Africa, South Korea,Spain, Sweden, Swizerland, Thailand, UAE, and the UK.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Philips Digital Diagnost Stationary radiographic system All Digital Diagnost systems with Eleva software version 3.0.x Class II When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction. When the operator manually rotates the image back, electronic side markers (if programmed) are rotated also and as a result may be placed in a wrong position inside the image. 360 Z-0900-2013
-
-