Event Detail

Event ID 63723
Product Type Devices
Status Ongoing
Recalling Firm Becton Dickinson & Co.
City Sparks
State MD
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-10-22
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution - USA (nationwide) to international customers in Belgium, India, Japan, South Korea, Thailand and Mexico

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
BD BBL Taxo XV Factor Strips, Catalog number 231104, in shelf pack vials, fifty strips/vial labeled in part ***Becton, Dickinson and Company, 7 Loveton Circle, Sparks, MD 21152 USA 800-638-8663, www.bd.com/ds*** Lot number 2089253 Class II In vitro diagnostic test kit was not manufactured according to specifications and may cause false negatives or misidentification of Hemophilus species present in patient specimens. 650 vials Z-0523-2013
-
-