| Event ID |
63809 |
| Product Type |
Devices |
| Status |
Ongoing |
| Recalling Firm |
Siemens Healthcare Diagnostics Inc |
| City |
Norwood |
| State |
MA |
| Country |
US |
| Voluntary/Mandated |
Voluntary: Firm Initiated |
| Recall Initiation Date |
2012-11-21 |
| Initial Firm Notification of Consignee or Public |
Letter |
| Distribution Pattern |
Worldwide Distribution - USA (nationwide) and internationally to Canada, Aruba, Brazil, Denmark, India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong, |
| Product Description |
Code Info |
Classification |
Reason for Recall |
Product Quantity |
Recall Number |
| Siemens Chek-Stix Positive and Negative Control Strips for Urine Chemistry
Catalog Number:1360
Chek-Stix® Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX® positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes |
Carton/Bottle Lot Number:
C9093061B P9093061B
C9094081B P9094081B
C0097091B P0097091B
C0100101C P0100101C
C0103042F P0103042F
C0106062B P0106062B
C91 01022D P91 01022D
CP9105052B P9105052B
C9108092A P9108092A
|
Class III |
When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25°C (77°F). |
7865 |
Z-0635-2013 |
| Siemens CLINITEK Atlas Positive and Negative Control Strips for Urine
Chemistry
Catalog Number: 5019
When used with the CLINITEK ATLAS® Automated urine Chemistry Analyzer and CLINITEK ATLAS® 10 reagent Pak, CLINITEK ATLAS® Positive and Negative Control Strips provide defined results for color, clarity, glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogen, nitrite, and leukocytes. Each CLINITEK ATLAS Positive Control Strip is a firm plastic strip to which are affixed seven separate analyte areas. |
Carton/Bottle Lot Number:
S9093061A P9093061A
S9094081A P9094081A
S0097091C P0097091C
S9093061A P9093061A
S9094081A P9094081A
S0097091C P0097091C
S0100101B P0100101B
S0103042A P0103042A
S0103042H P0103042H
S0106062A P0106062A
S9101022A P9101022A
S9101022B P9101022B
S9105052A P9105052A
|
Class III |
When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25°C (77°F). |
6466 |
Z-0636-2013 |
| Siemens Chek-Stix Combo Pak
Catalog Number: 1364
Chek-Stix® Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX® positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes |
Carton/Bottle Lot Number:
C9093061C P9093061C
C9094081J P9094081J
C0100101D P0100101D
C0097091E P0097091E
C9101022E P9101022E
C0099091D P0099091D
C0100101E P0100101E
C9101022H P9101022H
C9101022J P9101022J
C0103042B P0103042B
C0103042G P0103042G
C0106062D P0106062D
C9105052C P9105052C
|
Class III |
When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25°C (77°F). |
16711 |
Z-0637-2013 |
| Siemens Chek-Stix Positive and Negative Control Strips for Urine Chemistry
(Distributed OUS Only)
Catalog Number: 1360L
Chek-Stix® Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX® positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes |
Carton/Bottle Lot Number:
C9094081C P9094081C
C0100101A P0100101A
C9101022C P9101022C
|
Class III |
When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25°C (77°F). |
700 |
Z-0638-2013 |
| Siemens Chek-Stix Combo Pak (Distributed OUS only)
Catalog Number: 1364M
Chek-Stix® Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX® positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes |
Carton/Bottle Lot Number:
CP0099091B P0099091B
C0100101G P0100101G
C0103042E P0103042E
|
Class III |
When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25°C (77°F). |
325 |
Z-0639-2013 |
| Siemens Chek-Stix Combo Pak
(Distributed OUS only)
Catalog Number: 1364N
Chek-Stix® Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX® positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes |
Carton/BottleLot Number:
C0099091A P0099091A
C0100101 H P0100101 H
C0103042C P0103042C
|
Class II |
When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25°C (77°F). |
1130 |
Z-0640-2013 |
| Siemens Chek-Stix Combo Pak
(Distributed OUS)
Catalog Number: 1364E
Chek-Stix® Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX® positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes |
Carton/Bottle Lot Number:
C0099091C P0099091C
C0100101F P0100101F
C9101022I P9101022I
C0103042D P0103042D
C0106062E P0106062E
|
Class III |
When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25°C (77°F). |
4642 |
Z-0641-2013 |