Enforcement Report - Week of January 2, 2013
-
Back to Previous Page
Event Detail
| Event ID | 63814 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Roche Molecular Systems, Inc. |
| City | Branchburg |
| State | NJ |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-10-11 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Distributed in the states of GA, CA, NC, and MN. |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| COBAS AmpliPrep Instrument (CAP) with AMPLILINK (AL) Software v.3.3.5 Automated sample processing instrument used with CAP/CA and CAP/CTM. Roche Molecular Systems, Inc. 1080 US Highway 202 South Branchburg, NJ 08876 USA Made in Switzerland | CAP 394436 to 394663 | Class II | Certain COBAS AmpliPrep instruments using AMPILINK software v.3.3.4 had the sample clot detection flagging feature disabled during quality control release resulting in samples that may not be flagged as having clots. | 7 units | Z-0614-2013 |
-