Event Detail

Event ID 63822
Product Type Devices
Status Ongoing
Recalling Firm Philips Medical Systems (Cleveland) Inc
City Cleveland
State OH
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-10-24
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution - USA including AK, AK, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, OA MA, MD MI, MN, MO, MS, NC, D, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV; Internationally to Canada and other countries..

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Philips Medical Systems, Brilliance Computed Tomography (CT) X-Ray Scanners, Big Bore, Model #728243; and Brilliance Computed Tomography (CT) X-Ray Scanners, Big Bore Oncology, Model #728244. Brilliance CT Big Bore - Model #728243 - Serial Numbers: 7006, 7007, 7008, 7009, 7013, 7014, 7015, 7016, 7017, 7018, 7019, 7020, 7021, 7022, 7023, 7024, 7025, 7026, 7027, 7028, 7029, 7030, 7031, 7032, 7033, 7034, 7035, 7036, 7037, 7038, 7039, 7040, 7041, 7042, 7043, 7044, 7045, 7046, 7047, 7048, 7049, 7050, 7051, 7053, 7054, 7055, 7056, 7057, 7058, 7059, 7060, 7061, 7062, 7063, 7064, 7065, 7066, 7067, 7068, 7069, 7070, 7071, 7072, 7073, 7074, 7075, 7076, 7077, 7078, 7079, 7080, 7081, 7082, 7083, 7084, 7085, 7086, 7087, 7088, 7089, 7090, 7091, 7092, 7093, 7094, 7095, 7096, 7097, 7098, 7099, 7100, 7101, 7102, 7103, 7104, 7105, 7106, 7107, 7108, 7109, 7110, 7111, 7112, 7113, 7114, 7115, 7116, 7117, 7118, 7119, 7120, 7122, 7123, 7124, 7125, 7126, 7127, 7128, 7129, 7130, 7131, 7132, 7133, 7134, 7135, 7136, 7137, 7138, 7139, 7140, 7141, 7142, 7143, 7144, 7145, 7146, 7147, 7148, 7149, 7150, 7151, 7152, 7153, 7154, 7156, 7157, 7158, 7159, 7160, 7161, 7162, 7163, 7164, 7165, 7166, 7167, 7168, 7169, 7170, 7171, 7173, 7174, 7175, 7177, 7178, 7179, 7181, 7183, 7184, 7185, 7186, 7187, 7188, 7189, 7191, 7192, 7193, 7194, 7195, 7196, 7197, 7198, 7199, 7201, 7202, 7203, 7204, 7205, 7206, 7207, 7208, 7209, 7210, 7211, 7212, 7213, 7214, 7215, 7216, 7217, 7218, 7219, 7220, 7221, 7222, 7223, 7224, 7225, 7226, 7227, 7228, 7229, 7230, 7231, 7232, 7233, 7234, 7235, 7236, 7237, 7238, 7239, 7240, 7241, 7242, 7243, 7245, 7246, 7247, 7248, 7249, 7250, 7251, 7253, 7254, 7255, 7256, 7257, 7258, 7259, 7260, 7261, 7262, 7263, 7265, 7266, 7267, 7269, 7270, 7272, 7273, 7274, 7275, 7276, 7277, 7278, 7279, 7280, 7281, 7282, 7283, 7284, 7285, 7286, 7287, 7288, 7289, 7290, 7291, 7292, 7293, 7294, 7295, 7296, 7297, 7299, 7300, 7302, 7304, 7305, 7306, 7307, 7308, 7309, 7310, 7311, 7312, 7313, 7314, 7315, 7316, 7317, 7319, 7320, 7321, 7322, 7323, 7324, 7325, 7326, 7327, 7328, 7329, 7330, 7331, 7332, 7333, 7334, 7335, 7336, 7337, 7338, 7339, 7340, 7341, 7342, 7343, 7344, 7345, 7346, 7347, 7348, 7349, 7350, 7351, 7352, 7353, 7354, 7355, 7356, 7357, 7358, 7359, 7360, 7361, 7362, 7363, 7364, 7365, 7366, 7367, 7368, 7369, 7370, 7371, 7372, 7373, 7374, 7376, 7377, 7378, 7379, 7381, 7382, 7385, 7387, 7388, 7389, 7391, 7392, 7393, 7394, 7395, 7397, 7398, 7399, 7400, 7401, 7403, 7404, 7405, 7406, 7409, 7410, 7411, 7412, 7413, 7414, 7415, 7416, 7417, 7418, 7419, 7420, 7421, 7422, 7423, 7424, 7425, 7426, 7428, 7429, 7431, 7432, 7433, 7434, 7435, 7436, 7437, 7438, 7440, 7441, 7442, 7443, 7444, 7445, 7446, 7447, 7448, 7449, 7450, 7451, 7454, 7455, 7456, 7457, 7459, 7462, 7463, 7464, 7465, 7466, 7467, 7468, 7469, 7470, 7471, 7472, 7473, 7474, 7475, 7476, 7477, 7478, 7479, 7480, 7481, 7482, 7483, 7484, 7485, 7487, 7488, 7489, 7490, 7491, 7492, 7493, 7494, 7495, 7496, 7497, 7498, 7499, 7500, 7501, 7502, 7503, 7504, 7505, 7506, 7507, 7508, 7509, 7511, 7512, 7513, 7515, 7516, 7517, 7518, 7519, 7521, 7522, 7523, 7524, 7526, 7527, 7528, 7530, 7531, 7532, 7534, 7535, 7537, 7539, 7540, 7542, 7544, 7545, 7547, 7548, 7549, 7550, 7551, 7552, 7553, 7554, 7555, 7556, 7558, 7559, 7560, 7561, 7562, 7564, 7565, 7566, 7568, 7569, 7573, 7574, 7577, 7578, 7579, 7580, 7581, 7582, 7584, 7585, 7588, 7589, 7590, 7591, 7592, 7593, 7596, 7597, 7598, 7599, 7601, 7602, 7603, 7605, 7608, 7609, 7611, 7612, 7613, 7615, 7616, 7617, 7618, 7620, 7622, 7623, 7624, 7626, 7627, 7628, 7629, 7630, 7634, 7635, 7637, 7638, 7639, 7640, 7641, 7642, 7643, 7644, 7645, 7646, 7647, 7648, 7649, 7650, 7652, 7655, 7656, 7657, 7658, 7659, 7661, 7662, 7663, 7665, 7666, 7667, 7671, 7672, 7673, 7674, 7675, 7677, 7678, 7679, 7680, 7681, 7682, 7684, 7685, 7686, 7687, 7689, 7690, 7691, 7693, 7695, 7696, 7698, 7699, 7700, 7701, 7702, 7703, 7704, 7705, 7706, 7707, 7709, 7715, 7716, 7718, 7720, 7724, 7725, 7727, 7729, 7730, 7731, 7735, 7736, 7737, 7739, 7740, 7741, 7743, 7744, 7745, 7749, 7750, 7751, 7752, 7753, 7754, 7755, 7757, 7758, 7759, 7761, 7762, 7763, 7765, 7768, 7769, 7770, 7774, 7775, 7776, 7777, 7779, 7782, 7783, 7784, 7785, 7786, 7787, 7788, 7791, 7793, 7794, 7795, 7796, 7797, 7799, 7800, 7801, 7802, 7803, 7806, 7807, 7808, 7810, 7811, 7812, 7814, 7815, 7818, 7820, 7823, 7824, 7827, 7828, 7833, 7834, 75000, 7536, 7541, 7688, 7595, 7572, 7636, 7268, 7614, 7190, 7529, 7600, 7402, 7710, 7571, 7654, 7252, 7632, 7453, 7244, 7486, 7533, 7557, 7538, 7607, 7708, 7052, 7660, 7407, 7384, 7576, 7318, 7427, 7390, 7546, 7668, 7651, 7587, 7697, 7200, 7172, 7121, 7375, 7396, 7408, 7586, 7452, 7514, 7683, 7631, 7510, 7380, 7543, 7430, 7621, 7301, 7528, 7383, 7155, 7520, 7439, 7610, 7619, 7180, 7386, 7525, 7570, 7271, 7694, 7653, 7575, 7458, 7676, 7583, 7182, 7176, 7669, and 7604. Class II Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Projection data set is used in the radiation therapy workflow, then this may result in unintended delivery of radiation to healthy tissue and/or failure to deliver radiation to the target lesion. 765 units Z-0880-2013
Philips Medical Systems, Extended Brilliance Workspace (EBW) X-Ray Scanners, Model #728260. A component of the Computed Tomography X-Ray System. The EBW Serial Numbers subject to software updates are as follows: 8222, 8235, 8270, 8838, 8841, 8862, 9014, 9019, 9127, 10331, 10348, 10402, 10705, 10749, 10827, 11002, 11362, 11365, 11588, 11597, 11652, 11693, 11704, 11724, 11725, 11828, 11857, 11861, 11945, 11949, 11950, 11951, 11952, 11953, 11954, 11955, 11956, 11957, and 11979. Class II Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Projection data set is used in the radiation therapy workflow, then this may result in unintended delivery of radiation to healthy tissue and/or failure to deliver radiation to the target lesion. 39 units Z-0881-2013
Philips Medical Systems, GEMINI LXL - Model #882412; GEMINI TF 16 - Model #882470 and Model #882473; GEMINI TF 64 - Model #882471, Positron Emission Tomography (PET) and X-Ray Computed Tomography (CT) scanning systems, diagnostic imaging systems for fixed or mobile installations that combine Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The affected GEMINI systems are identified with the following Serial Numbers: GEMINI LXL Model # 882412, Serial Numbers: 14702 and 14707; GEMINI TF 16 Model #882470, Serial Numbers: 7129, 7121, 7123, 7128, 7096, 7133, 7185, 7187, 7226, and 7220; GEMINI TF 16, Model #882473, Serial Numbers: 7150, and and 7237;GEMINI TF 64, Model #882471, Serial Numbers: 7103, 7126, 7132, 7158, 7160, 7164, 7175, 7178, 7179, 7218, 7229, 7221, 7245, 7550, 7255, and 7256. Class II Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Projection data set is used in the radiation therapy workflow, then this may result in unintended delivery of radiation to healthy tissue and/or failure to deliver radiation to the target lesion. 30 units Z-0882-2013
Philips Medical Systems, GEMINI TF Big Bore, Model #882476, Positron Emission Tomography (PET) and X-Ray Computed Tomography (CT) scanning system, a diagnostic imaging systems for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The affected GEMINI Big Bore system is identified with the following Serial Numbers: 9004, 9005, 9006, 9008 ,9009, 9010, 9011, 9012, 9013, 9014, 9014, 9016, 9017, 9018, 9019, 9020, 9021, 9022, 9023, 9201, 9202, 9203, 9204, 9205, 9205, 9206, 9207, 9208, 9209, 9210, 9213, 9214, 9215, 9216, 9217, 9218, 9219, 9220, and 9221. Class II Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Projection data set is used in the radiation therapy workflow, then this may result in unintended delivery of radiation to healthy tissue and/or failure to deliver radiation to the target lesion. 37 Z-0883-2013
-
-