Enforcement Report - Week of January 23, 2013
-
Back to Previous Page
Event Detail
| Event ID | 64007 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Terumo Cardiovascular Systems Corporation |
| City | Ann Arbor |
| State | MI |
| Country | US |
| Voluntary/Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2013-01-07 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Nationwide and worldwide: AL , AR , AZ , CA , CO , CT , FL , GA , HI , IA , IL , IN , KS , KY , LA , MA , ME , MI, MO , NC , ND , NV , NY , OH , OK , PA , SD , TN , TX , UT , VA , WA , and WI. Bahrain, BRAZIL, CANADA, CHILE, Honduras, Singapore, Taiwan, Trinidad and Tobago, and UNITED ARAB EMIRATES (UAE). |
Product Detail
| Product Description | Code Info | Classification | Reason for Recall | Product Quantity | Recall Number |
|---|---|---|---|---|---|
| Sarns flexible arterial cannula, 6.7 mm (20Fr) OD with 3/8 in. connector, suture ring, 9.5 in (24 cm) long The Sarns High Flow Aortic Arch Cannula and Sarns Flexible Arterial Cannulae are indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery | Lot number: 0677286 and Catalog number: 13010. | Class II | Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in the operative field, TCVS has decided to remove all affected lots that were manufactured with the affected connector part. | 6830 total units for all products | Z-0666-2013 |
| Sarns flexible arterial cannula, 7.3 mm (22 Fr) OD with 3/8in connector, suture ring, 9.5in (24 cm) long The Sarns High Flow Aortic Arch Cannula and Sarns Flexible Arterial Cannulae are indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery | Lot number: 0677284 and Catalog number: 13020 | Class II | Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in the operative field, TCVS has decided to remove all affected lots that were manufactured with the affected connector part. | 6830 total all products | Z-0667-2013 |
| Sarns flexible arterial cannula, 8.0 mm (24 Fr) with 3/8in connector, suture ring, 9.5in (24 cm) long The Sarns High Flow Aortic Arch Cannula and Sarns Flexible Arterial Cannulae are indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery | Lot number: 067728 and 0681074 and Catalog number: 13030 | Class II | Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in the operative field, TCVS has decided to remove all affected lots that were manufactured with the affected connector part. | 6830 total all products | Z-0668-2013 |
| Sarns high-flow aortic arch cannula, 5.2 mm (16 Fr) OD with 3/8in connector, 7.5in (19 cm) long The Sarns High Flow Aortic Arch Cannula and Sarns Flexible Arterial Cannulae are indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery | Lot number: 0677294, 0680693, 0682187 and Catalog number: 12315 | Class II | Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in the operative field, TCVS has decided to remove all affected lots that were manufactured with the affected connector part. | 6830 total all products | Z-0669-2013 |
| Sarns high-flow aortic arch cannula, 6.5 mm (20 Fr) OD with 3/8in connector, 11in (28 cm) long The Sarns High Flow Aortic Arch Cannula and Sarns Flexible Arterial Cannulae are indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery | Lot number: 0680122, 0680751and Catalog number: 12325 | Class II | Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in the operative field, TCVS has decided to remove all affected lots that were manufactured with the affected connector part. | 6830 total all products | Z-0670-2013 |
| Sarns high-flow aortic arch cannula, 6.5 mm (20 Fr) OD with 3/8in connector, 11in (28 cm) long, with X coating surface The Sarns High Flow Aortic Arch Cannula and Sarns Flexible Arterial Cannulae are indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery | Lot number: 0677299 and Catalog number: 12325X | Class II | Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in the operative field, TCVS has decided to remove all affected lots that were manufactured with the affected connector part. | 6830 total all products | Z-0671-2013 |
-