Event Detail

Event ID 64307
Product Type Devices
Status Ongoing
Recalling Firm Animas Corporation
City West Chester
State PA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-12-17
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution-USA (nationwide) including Washington, DC, Puerto Rico, US Virgin Island and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY, and the countries of Austria, Australia, Bermuda, Canada, New Zealand, Czech Republic, Italy, France, Finland, Hungary, Israel, United Kingdom, Spain, Sweden, Norway, Denmark, Germany, Mexico and Ireland.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Animas(R) 2020, Animas(R) IR 1250 (1200 plus), Animas(R) IR 1200 Insulin Infusion Pumps Indicated for continuous subcutaneous infusion of insulin. All Animas 2020, IR1250 and IR1200 Pump Models - Part Numbers 101370-00, 101371-00, 101372-00, 101374-00, 101375-00, 100380-00, 100381-00, 100382-00, 100384-00, 100385-00, 100450-00, 100210-00, 100211-00, 100212-00, 100250-00, 100251-00, 100252-00, 100200-00, 100201-00, 100202-00, 100170-00, 100171-00 and 100172-00. Class III Due to an internal calendar date limit (December 31, 2015) these pumps will cease to operate on January 1, 2016, and will result in a call service alarm that is displayed on the pump display. 37,998 Z-0836-2013
-
-