Event Detail

Event ID 64579
Product Type Devices
Status Ongoing
Recalling Firm Biosense Webster, Inc.
City Irwindale
State CA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-03-05
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution - USA (Nationwide) and countries of: DE, NL, CH, FR, GB, DK.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
20-Pole Eco Cable, Catalog #EM-5050-060 Product Usage: The 20-Pole Eco Cable is intended to be used as an interface cable between the LASSO NAV Eco Catheter and CARTO 3 System. Impacted serial Numbers: 120, 132, 134,136, 151, 342, 350, 354, 356, 363, 368, 373, 379, 380, 384, 385, 402, 405, 424, 919, 953, 954, 960, 963, 968, 970, 977, 980, 993, 998, 1001, 1123, 1148, 1378, 1596, 1729, 1745, 1747, 1749, 1750, 1751, 1753, 1754, 1758, 1759, 1760, 1761, 1797, 1799, 1802, 1803, 1804, 1805. Lot Numbers: P1210, P1220, P1221, P1231. Class II Biosense is recalling the 20-Pole Eco Cable because it was incorrectly calibrated. When a catheter is connected to an affected to an affected 20-Pole Eco Cable, the catheter will not be recognized by the impacted 54 cable serial numbers within the field. 54 units Z-1023-2013
-
-