Event Detail

Event ID 64662
Product Type Devices
Status Ongoing
Recalling Firm Siemens Medical Solutions USA, Inc
City Malvern
State PA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-03-08
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Nationwide Distribution including AL, CA, CO, FL, IA, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, ND, NJ, NY, OH, PA, SC, TX, UT, VA, WA, and WI and to Puerto Rico.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Artis Systems with Software Artis VC20x/VC21A/VD10x that have a DSA license installed. Angiographic x-ray system Model numbers 10094135, 10094137, 10094139, 10094141 and 10280959. Class II The firm has become aware of a potential issue on Artis systems with the software Artis VC20x/VC21A/VD10x that have a DSA license installed. Under certain preconditions, executing roadmap OGP (organ program) in the DSA overlay mode may lead to imprecise registration of the DSA mask image with the roadmap image displayed on the live monitor. 81 Z-0997-2013
-
-