Event Detail

Event ID 64693
Product Type Devices
Status Ongoing
Recalling Firm Toshiba American Medical Systems Inc
City Tustin
State CA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-12-11
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Nationwide Distribution.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Aplio 500/400/300 Diagnostic Ultrasound System (TUS-500/400/300). Device Listing Number: Dl53118. Diagnostic Ultrasound system. Item numbers: TUS-A300/Wl and TUS-A500/Wl Class II Toshiba Medical Systems has received reports of startup issues with the Toshiba TUS-500 and TUS-300 ultrasound systems. The startup issue is predominant when attempting to bring the system up from standby mode when the system is used remotely. It has been determined that this may potentially cause damage to the system. 233 units Z-1104-2013
-
-