Event ID 64718
Product Type Devices
Status Ongoing
Recalling Firm Separation Technology, Inc.
City Sanford
State FL
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-03-13
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution-USA (nationwide) including the states of PA, NC, IL, KS, and FL, and the countries Canada and Turkey.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
SLIDERPREP PLUS SLIDE PREPARATION CENTRIFUGE. Produce a monolayer of cells onto a glass slide from any fluid suspension. Serial numbers: 0110, 0111, 0112, 0210, 0211, 0212, 0310, 0311, 0312, 0410, 0411, 0412, 0511, 0512, 0611, 0710, 0711, 0811, 0812, 0910, 0911, 0912, 1011, 1012, 1112, 1210, 1212. Class II Separation Technology, Inc. recalled SlidePrep Plus Cytology Centrifuges manufactured prior to January 1, 2013 for having the potential to start up unexpectedly when encountering an electrostatic discharge. 102 Z-1367-2013
-
-