Event Detail

Event ID 64723
Product Type Devices
Status Ongoing
Recalling Firm Horiba Instruments, Inc dba Horiba Medical
City Irvine
State CA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-03-20
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Nationwide Distribution.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
HORIBA ABX PENTRA 400 I.S.E. Module, Model Numbers: P400ISE110EN02 and P400ISE110US02, All Serial Numbers, Software Versions 5.0.7 and Below. Used to measure absorbance and ion electrodes. Model Numbers: P400ISE110EN02 and P400ISE110US02. All Serial Numbers. Versions 5.0.7 and below Class II HORIBA Medical is informing all ABX PENTRA 400 Analyzer with I.S.E. Module customers that there are no Linearity Limit Flags for the I.S.E. assays currently programmed into the ABX PENTRA 400 Analyzer software. All other assays run on this analyzer do have Linearity Limit Flags to alert the user to values that fall below or above the validated linearity range for each assay. 905 units Z-1141-2013
-
-