• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of May 29, 2013

  • Print
  • Share
  • E-mail
-

Event Detail

Event ID 64850
Product Type Devices
Status Ongoing
Recalling Firm Roche Diagnostics Operations, Inc.
City Indianapolis
State IN
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-04-04
Initial Firm Notification of
Consignee or Public
Distribution Pattern USA Nationwide Distribution

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Cobas® 8000 modular analyzer series software version 03-01, 02, 03 and 04, impacting the cobas e 602 module only. Product Usage: Fully automated immunoassay analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids. This analyzer is optimized for high throughput workloads in the professional environment using a combination of ion selective electrodes and photometric analysis modules Part Number 05990378001 Serial Numbers 1017-06, 1021-03, 1021-05, 1142-09, 1145-10, 1140-01, 1125-07, 1128-02, 1146-01, 1146-02, 1146-03, 1171-10, 1169-06, 1169-08, 1169-09, 1172-09, 1156-01, 1156-02, 1156-09, 1160-04, 1160-06, 1016-05, 1017-01, 1139-08, 1139-10, 1140-07, 1140-08, 1140-10, 1156-06, 1156-07, 1158-09, 1160-01, 1160-05, 1173-01, 1139-06, 1139-07, 2374-11, 1127-08, 1138-08, 1138-10, 1139-01, 1156-08, 1156-10, 1173-10, 1156-05, 1158-10, 1139-09, 1131-08, 1132-01, 1132-02, 1171-07, 1171-08, 1171-09, 1147-05, 1149-01, 1124-05, 1124-06, 1132-03, 1132-04, 1147-05, 1149-01, 1128-05, 1128-06, 1139-02, 1139-03, 1139-04, 1139-05, 1146-05, 1146-06, 1147-04, 1156-04, 1132-05, 1132-10, 1137-02, 1147-05, 1149-01, 1128-05, 1128-06, 1149-07, 1149-08, and 1167-02. Class II When using the cobas e 602 module, Diluent Multiassay may be used for autodilution of certain assays instead of Diluent Universal. This may cause incorrect results under specific circumstances if all of the following conditions are met simultaneously: More than a total of 2000 reagent kits or diluents have been registered on the cobas e 602 modules within one system core, and two different diluents are on board the module in parallel (i.e., Diluent Universal and Diluent MultiAssay), and Diluent Universal previously registered correctly is now being registered as Diluent MultiAssay, or vice versa. 72 units Z-1368-2013
-
-