Event Detail

Event ID 64985
Product Type Devices
Status Ongoing
Recalling Firm SpineFrontier, Inc.
City Beverly
State MA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-11-09
Initial Firm Notification of
Consignee or Public
E-Mail
Distribution Pattern US Distribution including Puerto Rico and the states of CO, KS, MA and TX.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Drill Guide - Chameleon FacetFuse Fixation System, Model: 11-00028 Rev. A FacetFuse Drill Guides were designed to provide a concentric pathway for the drills and taps to access the bone while protecting the soft tissue from the cutting geometry of the drills/taps. In addition the drill guide is used as a reference to measure depth that the drills and taps are protruding into bone, which is measured by depth markings at the proximal end of the drills/taps relative to the proximal end of the drill guide. These devices were used during implantation of spinal intervertebral body fusion devices intended for intervertebral body fusion of the spine of skeletally mature patients. 2520301, 2520307 Class II Prior to use in surgery, the Drill Guides were presenting noticeable degrees of denting and/or bending. As a result, drills and taps could not be directed down guides during surgery due to mechanical interference. 13 drill guides Z-1871-2013
-
-