Event Detail

Event ID 65042
Product Type Devices
Status Ongoing
Recalling Firm Zimmer, Inc.
City Warsaw
State IN
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-03-27
Initial Firm Notification of
Consignee or Public
E-Mail
Distribution Pattern Nationwide Distribution.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
NEXGEN Complete Knee Solution Monoblock Tibial Drill with stop 10.7 mm. Used to prepare cylindrical holes in the native tibia. Part No. 00-5887-052-00; lots 62181290 and 62183430 Class II Zimmer is initiating a lot specific recall of the Trabecular Metal" Tibia Stop Drill due to the potential of the outside diameter being oversized. As a result, there is a potential for the drill to not pass through the applicable drill guide. 69 units Z-1376-2013
-
-