• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of June 5, 2013

  • Print
  • Share
  • E-mail
-

Event Detail

Event ID 65067
Product Type Devices
Status Ongoing
Recalling Firm Polymer Technology Systems, Inc.
City Indianapolis
State IN
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-04-04
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution - USA Nationwide including Puerto Rico and countries of: UK, CHL, ENG, BEL, ITA, TKY, BEL, LIT, LTV, EGY, SPG, HOK, SWE, MEX, UAE, COR, ISR, MAL, AUS, POR, POL, SLN and STQ.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
PTS PANELS, Glucose Test Strips for use with CardioChek Brand Analyzers Product Usage: used by healthcare professionals and individuals with diabetes to measure glucose in whole blood U1117 Class II Internal investigation demonstrated under-recovery for glucose when tested against a reference method. This could result in a low to moderate health risk. 7048 total, 3499 in USA Z-1403-2013
-
-