Event Detail

Event ID 65147
Product Type Devices
Status Ongoing
Recalling Firm Siemens Healthcare Diagnostics, Inc.
City Newark
State DE
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-05-10
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution: USA (nationwide) and internationally to Canada

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
syngo(R) Lab Data Manager - Version VA11B and VA12A Systems Securely consolidates data from multiple lab instruments, the LIS and Siemens StreamLAB(R) Automation Solutions. It provides autoverification, integrated QC capabilities, browser-based connectivity supporting up to 25 concurrent users, remote services to help keep systems running optimally and also has audit trail capabilities. Siemens Material Number 10800057 (Version VA11B) and 10803188 (Version VA12A) Class II Under certain conditions the system may not perform as intended, causing the release of results to the Laboratory Information System that should have been held for manual review due to auto-verification rules or the delay/omission of result transmission to the LIS. 262 Z-0433-2014
-
-