• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of July 10, 2013

  • Print
  • Share
  • E-mail
-

Event Detail

Event ID 65159
Product Type Devices
Status Ongoing
Recalling Firm Maquet Cardiovascular Us Sales, Llc
City Wayne
State NJ
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-04-26
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution - USA including Arizona, Georgia, North Dakota, South Carolina, Texas and Tennessee, Internationally to United Arab Emirates, Austria, Brazil, Canada, Switzerland, China, Germany, Spain, France, United Kingdom, Italy, Japan, Singapore, Thailand, Turkey, Serbia, and South Africa.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
MAQUET Getinge Group Mobile Operating Table YUNO OTN GA 1433.01 EN 05. The YUNO OTN is used to support and position patients for surgical procedures immediately before, during and after the surgical phase. The table top is radio translucent and enables intraoperative use of X-ray equipment. YUNO OTN Mobile Operating Table: 1433.01F0 Extension Device: 1433.62A1 Class II MAQUET has identified a potential issue with the extension device handle (part number 1433.62A1) and will make enhancements to the YUNO table (part number 1433.01F0). Identification of the Extension device issue: By fully actuating the lever of the extension device handle, the operator's skin can enter a gap between the handle lever and a recess in the handle bar. When the lever is released, skin which might be within the gap could be pinched or cut. 13 Tables; 24 Extension devices Z-1622-2013
-
-