Event Detail

Event ID 65214
Product Type Devices
Status Ongoing
Recalling Firm Sysmex America, Inc.
City Mundelein
State IL
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-03-31
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and the states of AL, AR, CA, CO, CT, FL, GA, HI, IA, IS, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA and WI and the country of Canada.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
The UF-1000i is a fully automated urine particle analyzer intended for vitro diagnostic use in urinalysis. The instrument can perform screening for abnormal samples with a high degree of accuracy, so it increases automation and efficiency in laboratories. Software versions 1.0, 1.01, 1.1.1, 1.1.2 Class II The current Century Break patch did not contain the same setting for the Century Break year for both the Red Hat Linux and GUI files. Due to the mismatch in file settings, the software could not properly calculate a purge date for the database files based on the customer data retention setting in the application. This resulted in the database files meeting its maximum capacity and causing system performance issues. 214 units Z-0949-2014
-
-