Event Detail

Event ID 65248
Product Type Devices
Status Ongoing
Recalling Firm Endologix Inc
City Irvine
State CA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-05-13
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution - USA Nationwide in the states of: NY, NH, IN, MI, FL and NJ and in the countries of: Europe and Latin America.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Brand Name: AFX" Introducer System, Model Number S17-45. Lot Numbers: 1079840, 1079843, 1079844, 1079845. Product Usage: The AFX Introducer System is intended for use to facilitate the introduction of catheters and other medical devices into the vasculature and to minimize blood loss associated with such introduction. Model Number S17-45. Lot Numbers: 1079840, 1079843, 1079844, 1079845 Class I Endologix, Inc. initiated this voluntary recall of certain lots of Endologix, Inc. due to reports of dilator breakage during clinical procedures. 53 Z-1436-2013
-
-