Event ID 65605
Product Type Devices
Status Ongoing
Recalling Firm Siemens Medical Solutions USA, Inc
City Malvern
State PA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-05-02
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern US Distribution including the states of IA, SD, NC, WI, OH, NE, TX, KS, ID, MI, GA, AR and IL.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Mobiletta Mira Mobile X-ray System Model number 10273100 Class II Siemens issued a Field Safety Alert about the potential risk to users when operating the Mobilett Mira. When a Mobilett Mira is regularly exposed to extensive mechanical strain (driven over high thresholds for doors or elevators), it is possible that one of its main wheel shafts could break because of material fatigue, resulting in the chassis falling on the ground on the affected wheel side. This imposes a potential risk of squeezing the toes of the user or another person. 39 Z-1875-2013
-
-