• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of August 7, 2013

  • Print
  • Share
  • E-mail
-

Event Detail

Event ID 65669
Product Type Devices
Status Ongoing
Recalling Firm Synthes USA HQ, Inc.
City West Chester
State PA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-10-12
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern US Distribution including the states of CA, MN, NC and TN.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Synthes Large External Fixation, MR Conditional, Tube to Tube Clamp, Part Number 390.007, Lot Number 6800599 Synthes Large External Fixation is intended for use to provide treatment for long bone and pelvic fractures that require external fixation. Part Number 390.007, Lot Number 6800599 Class II It was discovered internally that there was a discrepancy between the insert provided with the Tube to Tube Clamp and what was called for in the Bill of Materials. Following a detailed investigation, it was determined that the incorrect insert was used which indicates that the device is MR Safe, when it is actually MR Conditional. 8 Z-1816-2013
-
-