Event Detail

Event ID 65688
Product Type Devices
Status Terminated
Recalling Firm Topcon Medical Laser Systems, Inc
City Santa Clara
State CA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-03-27
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution-USA and the countries of China, Hong Kong, Brazil, France, Japan, Netherlands, Pakistan, Malaysia, United Kingdom, South Korea, Thailand, Dubai, and South Africa.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
PASCAL and PASCAL Streamline (532 nm.577 nm); Treats the patient's eye. PASCAL Standard 2R/G System v2.x.x and PASCAL Streamline 2R/G Software v2.x.x Serial numbers: United States 33057910 33060410 33063610 66078611 66084911 66085111 66085511 Out of United States OMC-179 11138509 11141209 11147209 11147809 11148910 11149010 11149210 11149710 . 11149810 11150110 11150210 11150410 11150610 11150810 11150910 11151010 1 151110 11151211 33054310 33059310 33059510 33062810" 66075211' 66076711 66077111 66078011 66079911 6608391'1 66084611 66084811 66085011 66085211 Class II When using the "Favorites" function for saved settings, there is potential for a mismatch between displayed power and the actual power output. This may result in over or under treatment. 7 US, 33 foreign Z-1774-2013
-
-