Event ID 65805
Product Type Devices
Status Ongoing
Recalling Firm Carestream Health, Inc.
City Rochester
State NY
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-07-26
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution - US (nationwide) including the states of New York, Massachusetts, Pennsylvania, Illinois, Texas, California and Louisiana., and the countries of Australia, Canada, Finland, Italy, Switzerland and United Kingdom.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
DRX Revolution Mobile X-Ray System, Model DRXR-1, Verona Street Rochester, NY 14608 Catalog Numbers 1019397, 1023415, 1023423 --- Made in USA by Carestream Health, Inc. 150 The device is designed to perform radiographic x-ray examinations on all pediatric and adult patients, in all patient treatment areas. Serial numbers: U.S. - 134,152,173,174,200,205,222,312,349,353,412, Worldwide Serial numbers: 469, 445, 306, 410, 453, 204, 483, 300, 171, 218 Class II During an audit, Carestream discovered a DRX-1 System that captured an image which displayed an artifact. This issue was discovered in manufacturing. It occurs when the DRX product console software is updated to v5.6b software and the detector is not calibrated before use. 23 units installed total (11 domestically & 12 internationally) Z-0090-2014
-
-