Event Detail

Event ID 65855
Product Type Devices
Status Ongoing
Recalling Firm Endogastric Solutions Inc
City Redmond
State WA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-06-06
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Distributed nationwide and Italy.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
EndoGastric Solutions EsophyX2 Device with Serosa Fastener and Accessories. Models R2001 and R2002. Indicated for the treatment of symptomatic chronic gastroesophageal reflux disease. UN-EXPIRED devices with Expiration dates (8/31/2013, 11/30/2013, 12/31/2013, and 1/31/2014) and following LOT numbers: 401132, 401147, 401152, 401156, 401159, 401167, 401207, 401214, 401219, 401220, 401225, 401227, 401230, 401233, 401236, 401246, 401248, 401251, 401257, 401258, 401259, 401267, 401269, 401271, 401274, and 401276. EXPIRED devices with Expired dates (1/31/2013, 2/28/2013, 3/31/2013 4/30/2013, 5/31/2013, 6/30/2013, and 7/31/2013) and following LOT numbers: 400923, 400927, 400934, 400937, 400940, 400943, 400949, 400961, 400963, 400973, 400976, 400979, 400980, 400981, 400984, 400986, 400992, 400994, 400997, 400998, 401004, 401005, 401009, 401012, 401016, 401017, 401024, 401026, 401029, 401056, 401058, 401059, 401060, 401061, 401062, 401063, 401065, 401072, 401074, 401076, 401079, 401084, 401085, 401092, 401094, 401107, 401114, 401115, 401121, 401129, and 401130. Class II Endogastric Solutions, Inc. has received a limited number of reports relating to the loss of tissue mold control when operating the R2001 or R2002 EsophyX2 device. In one case, surgical intervention was necessary to remove the device. 5192 units in the US and 41 units outside the US Z-1982-2013
-
-