Event Detail

Event ID 66024
Product Type Devices
Status Ongoing
Recalling Firm Villa Radiology Systems LLC
City Oxford
State CT
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-08-02
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern US Nationwide Distribution including the states of AZ, MI, IN, UT, CT, AL, NY, PA, IL, MD, CA, MA, IA, OH, MS and MO.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Juno DRF System. Mobile X-ray system These are general, multi-function x-ray systems capable of routine radiographs and general fluoroscopy procedures PN 709020 Class II It has been discovered that the Air Kerma rate information is inconsistent in Continuous Fluoro modes 30x30 (12"x12") and 15x15 (6"x6). This could result in a deviation of the displayed values of Air Kerma Rate and Cumulated Air Kerma from the actual values greater then +/-35% and then in a failure to comply with the requirements of 21 CFR 1020.32(k)(6 22 units within US Z-2160-2013
-
-