Enforcement Report - Week of October 2, 2013

  • Print
  • Share
  • E-mail
-

Event Detail

Event ID 66049
Product Type Devices
Status Ongoing
Recalling Firm Maquet Cardiovascular Us Sales, Llc
City Wayne
State NJ
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-08-15
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution, including Nationwide (US) and foreign countries.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
QUADROX iD HMOD 2030 Diffusion Membrane Oxygenator; QUADROX-i ADULT HMO 70000 / HMO 71000; QUADROX-iD Pediatric HMOD 30000. The QUADROX oxygenators are intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. QUADROX-iD Adult catalog number HMOD 70000, lot # 70081567 QUADROX-i Adult catalog number VKMO 70000, lot # 70082663 QUADROX-i Adult catalog number VKMO 71000, lot # 70083974 Class II The manufacturer has conducted a review of the results of product testing performed during the last twelve months and has identified specific lots of product which may not meet specifications for bond strength on the inlet and/or outlet connector. 1,182 units Z-2262-2013
-
-