Event Detail

Event ID 66105
Product Type Devices
Status Ongoing
Recalling Firm OPTI Medical Systems, Inc
City Roswell
State GA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-08-09
Initial Firm Notification of
Consignee or Public
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Distribution Pattern Worldwide distribution: US (nationwide) including states of: CA, MA, MN, and VA; and countries of: China, Guatemala, India, Iraq, South Korea, and Untied Arab Emirates.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Model BP7507 OPTI LION E-Plus Cassettes 25 Cassettes with Sample Probes, OPTIMedical Cassettes are consumable used in the OPTI LION Electrolyte Analyzer intended to be used for the measurement of sodium, potassium, chloride, ionized calcium and pH in samples of whole blood, serum, and plasma. Lots 252810, 315815, 319810, 319815, 321817, 367812, 370812 Class II OPTI LION E-Plus Cassettes were labeled with an incorrect barcode. The incorrect barcode could perfect the performance of one or more parameters as the product ages. Also a Customer complaint from a distributor reported that a customer was getting QC failures. 361 boxes of 25 cassettes each Z-0029-2014
-
-