Event Detail

Event ID 66155
Product Type Devices
Status Ongoing
Recalling Firm Cincinnati Sub-Zero Products Inc
City Cincinnati
State OH
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-09-03
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution - USA including AK, AR, AZ, CA, CO, DC, DE, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MT, , NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI & WV. Internationally to ARGENTINA, AUSTRALIA, AZERBAIJAN, BELGIUM, CANADA, CHILE, CHINA, COSTA RICA, CROATIA, GERMANY, HONDURAS, HONG KONG, INDIA, INDONESIA, ISRAEL, KOREA, KUWAIT, MEXICO, NEW ZEALAND, PERU, SAUDI ARABIA, SINGAPORE, SPAIN, TAHITI, THAILAND, THE NETHERLANDS, THE PHILIPPINES, UNITED ARAB EMERITES & UNITED KINGDOM.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Blanketrol III, Model Number 233, Cincinnati Sub-Zero Products, Inc., 12011 Mosteller Road, Cincinnati, OH 45241 Hyper-Hypothermia device used to lower or to raise a patients temperature and/or maintain a desired patient temperature through conductive heat transfer. Part number: 86107/06107/6107 (115V), Serial #s: 112-3-02021, 112-3-02102 through 124-3-02783, 074-3-00388, 081-3-00489, 081-3-00493, 082-3-00511, 091-3-00935, 094-3-01261, 094-3-01265, 094-3-01272, 094-3-01304, 094-3-01308, 101-3-01384, 093-3-01045, 093-3-01046, 093-3-01069, 084-3-00775, 073-3-00315, 091-3-00918, 091-3-00919, 073-3-00279, 094-3-01262, 071-3-00183, 092-3-00983, 094-3-01255, 064-3-00114, 072-3-00219, 072-3-00222, 072-3-00230, 072-3-00258, 081-3-00499, 081-3-00500, 064-3-00116, 064-3-00117, 092-3-01021, 073-3-00305, 073-3-00314, 082-3-22572, 082-3-22573, 101-3-01380, 073-3-00317, 074-3-00341, 091-3-00932, 094-3-01310, 101-3-01374, 082-3-00559, 092-3-00983, 094-3-01255, 094-3-01305, 062-3-00029, 093-3-01048, 094-3-01240, 102-3-01504 & 094-3-01243. Part number: 86207/6207 (230V) Serial #s: 112-3-01949, 114-3-02171 through 124-3-02764, 072-3-00247, 073-3-00266, 073-3-00282, 082-3-00564, 071-3-00132, 071-3-00152, 073-3-00309, 082-3-00533, 082-3-00534, 082-3-00565, 082-3-00568, 084-3-00768, 084-3-00770, 084-3-00777, 084-3-00778, 084-3-00779, 084-3-00780, 084-3-00781, 084-3-00787, 084-3-00788, 091-3-00836, 091-3-00877, 091-3-00878, 091-3-00902, 092-3-01007, 092-3-01008, 092-3-01011, 102-3-01423, 102-3-01426, 103-3-01592, 103-3-01593, 103-3-01594, 103-3-01595, 103-3-01601, 093-3-01072, 093-3-01073, 093-3-01076, 093-3-01078, 093-3-01084, 093-3-01085, 093-3-01086, 093-3-01087, 093-3-01088, 093-3-01104, 093-3-01105, 093-3-01106, 093-3-01107, 093-3-01108, 103-3-01573, 103-3-01574, 103-3-01575, 071-3-00178, 071-3-00180, 074-3-00393, 074-3-00394, 074-3-00395, 074-3-00396, 074-3-00404, 074-3-00405, 091-3-00904, 091-3-00905, 092-3-00939, 092-3-00948, 092-3-00949, 092-3-00951, 092-3-00962, 092-3-00963, 094-3-01178,094-3-01179, 094-3-01180, 094-3-01181, 094-3-01183, 084-3-00789, 084-3-00790, 084-3-00791, 084-3-00812, 084-3-00813, 084-3-00814, 084-3-00815 & 084-3-00816. Class II During a label review it was discovered that both the Blanketrol III Operations (57201-N) & Operations & Technical (56201-Q) manuals had verbiage added back into the manual during ECN M111-4260 (released 12/29/2011) which was previously removed for clarification of the intended use and subsequent 501(K) approval. 115V - 396 units; 230V - 385 units Z-2209-2013
-
-