Event Detail

Event ID 66239
Product Type Devices
Status Ongoing
Recalling Firm Roche Diagnostics Operations, Inc.
City Indianapolis
State IN
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-09-16
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Nationwide Distribution and Puerto Rico

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
COBAS INTEGRA 800 Analyzer with software 9864.C2 Cobas. COBAS Part Numbers 28122474001, 28122474692, 04559126001, 04589459970 corresponding to the Analyzer with or without closed tube sampling accessory A fully automated system for clinical chemistry analysis intended for the in vitro quantitative/qualitative determination of analytes in body fluids. software version 9864.C2 Class II Roche COBAS INTEGRA 800 system, Software Version 9864.C2 does not perform necessary ISE Service Actions. Whenever the ISE rack is removed, the ISE counters in the software are reset to zero and the corresponding Service Actions are not conducted automatically by the system. Additionally, certain maintenance activities that have to be performed manually by the operator are not requested on the software by the due date. This software issue ONLY affects systems performing ISE tests. 527 units Z-0168-2014
-
-