Event Detail

Event ID 66312
Product Type Devices
Status Ongoing
Recalling Firm Philips Medical Systems (Cleveland) Inc
City Cleveland
State OH
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-08-07
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution - US (nationwide) including FL, CO, KY, MS, PA, NC, NJ, TX and Puerto Rico and internationally to Italy.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
Philips MR Systems: Asset 0.5T, Apollo 0.5T, Infinion 1.5T, Eclipse/Polaris 1.5T, Panorama 0.6T 1.0T, Magnetic Resonance Imaging Systems intended for clinical/radiologic diagnostic imaging applications. M/N: 781220, 781230, 781301, 78129 and 78127. Codes: AM128, 1510, 1518, BA318, AM120, ZA178, ZA196, ZA107, ZA133, ZA172, ZA158, BA149 and ZA208. Class II There is a potential for water to collect in the vent pipe elbow in the magnet venting system. Water may freeze, blocking the venting system. Total 511 (Asset 0.5T,: 35 units Apollo 0.5T: 15 units, Infinion 1.5T: 116 units, Eclipse/Polaris 1.5T: 290 units, Panorama 0.6T: 55 units) Z-0033-2014
-
-