Event Detail

Event ID 66317
Product Type Devices
Status Ongoing
Recalling Firm Advanced Sterilization Products
City Irvine
State CA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-09-18
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Nationwide Distribution

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
STERRAD 100NX Cassettes The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices. Lot# 13F050, 13F051, 13F049 Class II Advanced Sterilization Products (ASP) is recalling the three lots (13F049, 13F050, 13F051) STERRAD 100NX Cassettes because ASP has determined that the leak indicator on the cassettes is effective for only 12 months of the cassette's labeled 15-month shelf-life. 2,995 units Z-0044-2014
-
-