Event Detail

Event ID 66947
Product Type Devices
Status Ongoing
Recalling Firm Bio-Rad Laboratories, Inc.
City Hercules
State CA
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-11-22
Initial Firm Notification of
Consignee or Public
Letter
Distribution Pattern Worldwide Distribution - USA (nationwide) and Internationally to Latin America, Asia Pacific, Canada, Australia, New Zealand and India.

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
D-10 Dual Program, which contains Update Kit D-10 A1c program diskette. United States; Bio-Rad Laboratories, Hercules, CA 94547 The Bio-Rad D-10 Hemoglobin Testing System provides an integrated method for the separation and determination of the relative percent of specific hemoglobins (e.g., A2, F, Ale} in whole blood. The separation is based on the principles of high performance liquid chromatography. Catalog number: 220-0201: Reorder Pack Lot No./Expiration date: 20030322 3/31/2014 20030337 3/31/2014 64000081 5/31/2014 64000331 5/31/2014 64000332 5/31/2014 64000888 08/31/2014 64001344 04/30/2014 64002239 11/30/2014 ( 64002240 08/3112014 64002241 11/30/2014 Class III D-10 A1c Dual Program Floppy was programmed incorrectly and cartridge injections counts are not being updated correctly when switching between the D-10 Dual Short Program and Extended Program on the D-10 Hemoglobin Testing System. 2906 kits (318 Kits in US; 2588 kits internationally) Z-0965-2014
-
-