• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of May 14, 2014

  • Print
  • Share
  • E-mail
-

Event Detail

Event ID 67871
Product Type Devices
Status Ongoing
Recalling Firm Med-Mizer, Inc.
City Batesville
State IN
Country US
Voluntary/Mandated Voluntary: Firm Initiated
Recall Initiation Date 2014-04-18
Initial Firm Notification of
Consignee or Public
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Distribution Pattern USA (nationwide)

Product Detail

Product Description Code Info Classification Reason for Recall Product Quantity Recall Number
RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 801, 802, 803, 803A: 80 " bed with 1, 2, or 3, motors and autocontour 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care. Model 801 : 50000-50018; Model 802n : 600000-600789 Model 803: 500000-500342; 803A 700000-701724 Class II All configurations of the SS Retractabed, Clinical Contour, made prior to June 7, 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holding the two support rails together when using the foot side rails and keeps them from separating from the frame. If the weld on the tube breaks, it can lead to bed collapse or the rollers coming out of the channel on the bed preventing bed high/low and contour functionality from occurring. 2980 Z-1543-2014
RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 799, 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care. Serial Number 2000 Class II All configurations of the SS Retractabed, Clinical Contour, made prior to June 7 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holding the two support rails together when using the foot side rails and keeps them from separating from the frame. If the weld on the tube breaks, it can lead to bed collapse or the rollers coming out of the channel on the bed preventing bed high/low and contour functionality from occurring. 1 Z-1542-2014
RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 762 ( 76 " bed w/ 2 motors); 763 (3 motors); 763 A ( 3 motors, autocontour) 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care. Model 762: 300000-300191 Model 763: 200000 Model 763A: 400000-40103 Class II All configurations of the SS Retractabed, Clinical Contour, made prior to June 7 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holding the two support rails together when using the foot side rails and keeps them from separating from the frame. If the weld on the tube breaks, it can lead to bed collapse or the rollers coming out of the channel on the bed preventing bed high/low and contour functionality from occurring. 191 Z-1541-2014
-
-