Product Detail

Product Description Philips MR Systems: Asset 0.5T, Apollo 0.5T, Infinion 1.5T, Eclipse/Polaris 1.5T, Panorama 0.6T 1.0T, Magnetic Resonance Imaging Systems intended for clinical/radiologic diagnostic imaging applications.
Recall Number Z-0033-2014
Classification Class II
Code Info M/N: 781220, 781230, 781301, 78129 and 78127. Codes: AM128, 1510, 1518, BA318, AM120, ZA178, ZA196, ZA107, ZA133, ZA172, ZA158, BA149 and ZA208.
Product Distributed Qty Total 511 (Asset 0.5T,: 35 units Apollo 0.5T: 15 units, Infinion 1.5T: 116 units, Eclipse/Polaris 1.5T: 290 units, Panorama 0.6T: 55 units)
Reason For Recall There is a potential for water to collect in the vent pipe elbow in the magnet venting system. Water may freeze, blocking the venting system.

Event Detail

Event Id 66312
Product Type Devices
Status Ongoing
Recalling Firm Philips Medical Systems (Cleveland) Inc
City Cleveland
State OH
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-08-07
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - US (nationwide) including FL, CO, KY, MS, PA, NC, NJ, TX and Puerto Rico and internationally to Italy.
-
-