Product Detail

Product Description STERRAD 100NX Cassettes The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.
Recall Number Z-0044-2014
Classification Class II
Code Info Lot# 13F050, 13F051, 13F049
Product Distributed Qty 2,995 units
Reason For Recall Advanced Sterilization Products (ASP) is recalling the three lots (13F049, 13F050, 13F051) STERRAD 100NX Cassettes because ASP has determined that the leak indicator on the cassettes is effective for only 12 months of the cassette's labeled 15-month shelf-life.

Event Detail

Event Id 66317
Product Type Devices
Status Ongoing
Recalling Firm Advanced Sterilization Products
City Irvine
State CA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-09-18
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Nationwide Distribution
-
-