Product Detail

Product Description Battery, 220V (Includes charger and connecting cable) The Sarns" Centrifugal System is indicated for use in cardiopulmonary bypass procedures only.
Recall Number Z-0050-2013
Classification Class II
Code Info Catalog number: 9491; and lot numbers: 1001-1187, 1101, 1193-1286, 2083-2085, 2018, 2288, 2289, 2447-2451, 2634, 2635, and 2884.
Product Distributed Qty 289 units
Reason For Recall Terumo Cardiovascular Systems (Terumo CVS) has received multiple reports since 2004 of various malfunctions of the battery backup system for the Sarns" Centrifugal System. The reports can be broadly categorized as follows: " Failure to charge " Failure to hold charge " Failure to operate In the most severe case, a battery malfunction could result in an unexpected loss of arterial flow. In this case, if the user doesn't clamp the venous line quickly enough, the patient could experience hypovolemia. In the event of a backflow condition, air could become entrained in the arterial line, which if not detected, could be passed to the patient.

Event Detail

Event Id 63079
Product Type Devices
Status Ongoing
Recalling Firm Terumo Cardiovascular Systems Corporation
City Ann Arbor
State MI
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-09-25
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide distribution: USA including AK, AL, AR, AZ, CA, CO, CT, DC
-
-