Enforcement Report - Week of October 24, 2012
-
Back to Previous Page
Product Detail
| Product Description | "***Electric Systems Foot Control****Rx Only***" Product Usage: Cutting and shaping bone including bones of the spine and cranium. |
|---|---|
| Recall Number | Z-0074-2013 |
| Classification | Class II |
| Code Info | Serial Numbers: E31305311301, E31305311302, E31305311303, E31305311304, E31305311401, E31305311402, E31305311403, E31305311404, E31305311405, E32305311201, E32305311202, E33305311501, E33305311502, E33305311504, and E33305311505 |
| Product Distributed Qty | 16 units |
| Reason For Recall | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling the foot control/foot pedal due to the Foot Control Cable lacking the shield connection from the cable to the connector body. |
Event Detail
| Event Id | 63086 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | The Anspach Effort, Inc. |
| City | Palm Beach Gardens |
| State | FL | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-01-09 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide Distribution - US (nationwide) and the countries of: Italy, Japan, Switzerland, Turkey and Taiwan. |
-