Enforcement Report - Week of October 24, 2012
-
Back to Previous Page
Product Detail
| Product Description | "***Electric Systems Foot Control with Direction and Irrigation***Rx Only***" Product Usage: Cutting and shaping bone including bones of the spine and cranium. |
|---|---|
| Recall Number | Z-0076-2013 |
| Classification | Class II |
| Code Info | Serial Numbers: E3230327501, E32305327502, E32305327503, E32305327504, E32305327505, E32305327601, E32305327602, E32305327603, E32305327604, and E32305327605 |
| Product Distributed Qty | 10 units |
| Reason For Recall | The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling the foot control/foot pedal due to the Foot Control Cable lacking the shield connection from the cable to the connector body. |
Event Detail
| Event Id | 63086 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | The Anspach Effort, Inc. |
| City | Palm Beach Gardens |
| State | FL | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-01-09 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide Distribution - US (nationwide) and the countries of: Italy, Japan, Switzerland, Turkey and Taiwan. |
-