• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of November 6, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Fujinon/Fujifilm Ultrasonic Endoscope Intended to provide ultrasonic images of submucosal and peripheral organs of the upper gastrointestinal tract for observation, diagnosis and endoscopic treatment.
Recall Number Z-0112-2014
Classification Class II
Code Info Model Number EG-530UT2. Serial Numbers: NU038A018, NU038A019, NU038A022, NU038A027, NU038A028 and NU038A029
Product Distributed Qty 12 Units
Reason For Recall A limited number of endoscopes were shipped to FMSU-ESD with a manufacturing defect which included missing set screws used to more securely attach the Forceps Inlet (biopsy inlet port) to the endoscope.

Event Detail

Event Id 66144
Product Type Devices
Status Ongoing
Recalling Firm Fujinon Inc
City Wayne
State NJ
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-07-31
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide distribution: US (nationwide) including states of: NJ, PA and VA; and country of Canada.
-
-