• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of November 13, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Dual Luer Lock Cap The device is intended for use as a cap for male or female Luer ports on medical devices such as manifolds, stopcocks or sets.
Recall Number Z-0143-2014
Classification Class II
Code Info 10043, 10044
Product Distributed Qty 628,992 units
Reason For Recall The firm is recalling lots 10043 and 10044 due to loose particulate matter found in the packaging. Particulate matter entering the fluid path from the Luer lock caps may result in embolic events.

Event Detail

Event Id 66249
Product Type Devices
Status Ongoing
Recalling Firm Baxter Healthcare Corp.
City Deerfield
State IL
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-09-10
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - USA (nationwide) and the country of Canada.
-
-